Loading Events

Infrared thermometer calibration

EPTIS ID: 1394698
PT scheme / program name: Thermodynamic calibration
PT item type: Temperature
Sample / item: infrared thermometer
Model / type: Minolta Land Cyclops 152
External provider for the calibration of the PT item:  an ISO/IEC 17025 accredited laboratory or NMI, used by AG Metrology (PT provider) to determine the assigned value

Measurement points and CMC from the calibration laboratory’s accreditation schedule:

Measurement Points CMC
600 °C 0.90 °C
800 °C 0.90 °C
1000 °C 1.0 °C
1200 °C 1.0 °C
1400 °C  1.5 °C

The uncertainty actually used depends on the istrument and a stability  contribution, finalised each round. The laboratory only calibrates; AG Metrology alone determines the assigned value and is responsible for the PT scheme.

Note: Indications regarding organizational methods, presentation and sending of results will be reported in the PT participation protocol

Scheduled start date for PT: even-numbered years in July

Payment: bank transfer

For further information

CONTACT US
Go to Top