5½ digit multifunction calibrator, AC calibration

PT scheme / program name: Electrical calibration
PT item type: AC/DC Voltage, AC/DC Current, Resistance
Sample / item: 5½ digit multifunction calibrator AC
Model: Meatest M143
Reference Laboratory: AG Metrology S.r.l.
Measurement point and Reference Laboratory CMC

Type of Test CMC Measurement Point
ACV up to 1000 V from 40 Hz up to 10 kHz
10 mV
(40, 400) Hz 0.026 % rdg + 15 µV 10 mV
10 kHz 0.041 % rdg + 13 µV 10 mV
100 mV
(40, 400) Hz 0.026 % rdg + 15 µV 100 mV
10 kHz 0.041 % rdg + 13 µV 100 mV
1 V
(40, 400) Hz 0.021 % rdg + 0.11 mV  1.0 V
10 kHz 0.030 % rdg + 53 µV 1.0 V
10 V
(40, 400) Hz 0.021 % rdg + 1.1 mV 10 V
10 kHz 0.029 % rdg + 0.51 mV 10 V
100 V
(40, 400) Hz 0.021 % rdg + 11 mV 100 V
10 kHz 0.021 % rdg + 5.1 mV 100 V
1000 V
(40, 400) Hz 0.026 % rdg + 0.15 V 900 V
10 kHz 0.031 % rdg + 0.13 V 900 V
ACI up to 9 A from 20 Hz up to 5 kHz
100 µA
(0.05, 1) kHz 0.041 % rdg + 33 nA (10, 90) µA
5 kHz 0.041 % rdg + 33 nA 90 µA
1 mA
(0.05, 1, 5) kHz 0.048 % rdg + 0.24 µA 0.9 mA
10 mA
(0.05, 1, 5) kHz 0.048 % rdg + 2.7 µA 9 mA
100 mA
(0.05, 1, 5) kHz 0.048 % rdg + 27 µA 90 mA
1 A
(0.05, 1, 5) kHz 0.17 % rdg + 0.84 mA 0.9 A
10 A
(0.05, 1, 5) kHz 0.36 % rdg + 15 mA 9 A

Note: Indications regarding organizational methods, presentation and sending of results will be reported in the PT participation protocol