PT DMM: Calibration 6½ digit DMM, AC
Item: DMM Agilent 34401A 6½ digit AC
Model: Agilent 34401A
Reference Laboratory: ISO IEC 17025 accreditated laboratory or NMI
Measurement point and Reference Laboratory CMC
Type of Test | CMC | Measurement Point |
---|---|---|
ACV up to 675 V from 50 Hz up to 100 kHz | ||
100 mV | ||
1 kHz | 0.034% rdg + 18 µV | 10 mV, 90 mV |
50 Hz | 0.034% rdg + 18 µV | 90 mV |
20 kHz | 0.046% rdg + 27 µV | 90 mV |
50 kHz, 100 kHz | 0.18% rdg + 46 µV | 90 mV |
1 V | ||
50 Hz, 1 kHz | 0.029% rdg + 0.11 mV | 0.9 V |
20 kHz | 0.030% rdg + 0.11 mV | 0.9 V |
50 kHz, 100 kHz | 0.12% rdg + 0.41 mV | 0.9 V |
10 V | ||
50 Hz, 1 kHz | 0.029% rdg + 1.1 mV | 1 V, 5V, 9V |
20 kHz | 0.029% rdg + 1.1 mV | 9 V |
50 kHz, 100 kHz | 0.12% rdg + 4.1 mV | 9 V |
100 V | ||
50 Hz, 1 kHz, 20 kHz | 0.025% rdg + 11 mV | 90 V |
50 kHz, 100 kHz | 0.11% rdg + 41 mV | 90 V |
1000 V | ||
50 Hz | 0.037% rdg + 0.16 V | 675 V |
5 kHz, 10 kHz | 0.035% rdg + 0.25 V | 675 V |
ACI up to 0,9 A from 20 Hz up to 5 kHz | ||
1 A | ||
50 Hz, 1 kHz | 0.048% rdg + 27 µA | 0.1 A |
20 Hz, 50 Hz, 1 kHz, 5 kHz | 0.17% rdg + 0.84 mA | 0.9 A |
Note: Indications regarding organizational methods, presentation and sending of results will be reported in the PT participation protocol
Scheduled start date for PT: April 2023
Payment: by bank transfer