Loading Events
This event has passed.

PT DMM: Calibration 6½ digit DMM, AC

Item: DMM Agilent 34401A 6½ digit AC
Model: Agilent 34401A
Reference Laboratory: ISO IEC 17025 accreditated laboratory or NMI
Measurement point and Reference Laboratory CMC

Type of Test CMC Measurement Point
ACV up to 675 V from 50 Hz up to 100 kHz
100 mV
1 kHz 0.034% rdg + 18 µV 10 mV, 90 mV
50 Hz 0.034% rdg + 18 µV 90 mV
20 kHz 0.046% rdg + 27 µV 90 mV
50 kHz, 100 kHz 0.18% rdg + 46 µV 90 mV
1 V
50 Hz, 1 kHz 0.029% rdg + 0.11 mV 0.9 V
20 kHz 0.030% rdg + 0.11 mV 0.9 V
50 kHz, 100 kHz 0.12% rdg + 0.41 mV 0.9 V
10 V
50 Hz, 1 kHz 0.029% rdg + 1.1 mV 1 V, 5V, 9V
20 kHz 0.029% rdg + 1.1 mV 9 V
50 kHz, 100 kHz 0.12% rdg + 4.1 mV 9 V
100 V
50 Hz, 1 kHz, 20 kHz 0.025% rdg + 11 mV 90 V
50 kHz, 100 kHz 0.11% rdg + 41 mV 90 V
1000 V
50 Hz 0.037% rdg + 0.16 V 675 V
5 kHz, 10 kHz 0.035% rdg + 0.25 V 675 V
ACI up to 0,9 A from 20 Hz up to 5 kHz
1 A
50 Hz, 1 kHz 0.048% rdg + 27 µA 0.1 A
20 Hz, 50 Hz, 1 kHz, 5 kHz 0.17% rdg + 0.84 mA 0.9 A

Note: Indications regarding organizational methods, presentation and sending of results will be reported in the PT participation protocol